Identifier

etd-092911-134747

Abstract

"

Electroencephalogram (EEG), Electrocardiogram (ECG), and Electromyogram (EMG) bio-potential signals are commonly recorded in clinical practice. Typically, patients are connected to a bulky and mains-powered instrument, which reduces their mobility and creates discomfort. This limits the acquisition time, prevents the continuous monitoring of patients, and can affect the diagnosis of illness. Therefore, there is a great demand for low-power, small-size, and ambulatory bio-potential signal acquisition systems.

Recent work on instrumentation amplifier design for bio-potential signals can be broadly classified as using one or both of two popular techniques: In the first, an AC-coupled signal path with a MOS-Bipolar pseudo resistor is used to obtain a low-frequency cutoff that passes the signal of interest while rejecting large dc offsets. In the second, a chopper stabilization technique is designed to reduce 1/f noise at low frequencies. However, both of these existing techniques lack control of low-frequency cutoff.

This thesis presents the design of a mixed- signal integrated circuit (IC) prototype to provide complete, programmable analog signal conditioning and analog-to-digital conversion of an electrophysiologic signal. A front-end amplifier is designed with low input referred noise of 1 uVrms, and common mode rejection ratio 102 dB. A novel second order sigma-delta analog- to-digital converter (ADC) with a feedback integrator from the sigma-delta output is presented to program the low-frequency cutoff, and to enable wide input common mode range of ¡Ãƒâ€œ0.3 V. The overall system is implemented in Jazz Semiconductor 0.18 um CMOS technology with power consumption 5.8 mW from ¡Ãƒâ€œ0.9V power supplies. "

Publisher

Worcester Polytechnic Institute

Degree Name

PhD

Department

Electrical & Computer Engineering

Project Type

Dissertation

Date Accepted

2011-09-29

Accessibility

Unrestricted

Subjects

EEG, ECG, EMG, Sigma-Delta ADC.